| Q15.)
Does the InsiteIG fluorescence technology infringe on any U.S. patents?
Answers
Q1.) How can dissolved oxygen be measured
using fluorescence? BACK TO TOP
A very specific energy wavelength is transmitted to a ruthenium compound
immobilized in a sol-gel matrix. The ruthenium will absorb this energy,
changing the outer electron’s energy level. The electron will then
collapse back to it’s original energy state, emitting the energy
as a photon with a different specific wavelength. This is called fluorescing.
If the intensity of the transmitted wavelength is tightly controlled,
the amount of fluorescing is both predictable and repeatable.
If oxygen molecules are present the amount of fluorescing is reduced,
referred to as fluorescence quenching. By measuring the amount of quenching
it is possible to determine the amount of oxygen present.
Q2.) How long has this technology been
in use? BACK TO TOP
Fluorescence technology has been used to measure oxygen uptake in the
medical industry for over twenty years. Within the last five years the
technology has been adapted for use in measuring dissolved oxygen in biological
reactors. The key issues in making the technology viable in the water
treatment industry were the durability of the sensing element and cost.
Another important concern was packaging the system in a way that was easy
to use and required very little maintenance. The Insite units have accomplished
all of these objectives.
Q3.) What is the expected life of the
sensor? BACK TO TOP
Seven to ten years. During this time there are no consumables at all.
No spare parts, no recharging kits, no replacement films, and no membranes
or membrane cartridges.
Q4.) How often do the sensors need to
be calibrated? BACK TO TOP
We recommend that the calibration be checked at least once a year. The
sensor will drift less than one percent per year.
Q5.) Can the sensor be calibrated in
the field? BACK TO TOP
Yes, the sensor can be calibrated in the field. A simple calibration to
a reference takes about a minute.
Q6.) Are the sensors interchangeable, or do
the sensors have to remain with a specific analyzer?
BACK TO TOP
All Insite sensors can be moved at will. The microprocessor in the sensor
will “talk” to the microprocessor in the analyzer, identifying
itself. The analyzer will then configure itself to operate with that sensor.
Q7.) Will exposure to sunlight harm the
sensor or shorten the life of the sensor? BACK
TO TOP
No, the Insite sensor is completely unaffected by direct sunlight, unlike
some luminescent type sensors that can experience significant degradation
with as little as one hour of exposure to direct, or even indirect sunlight.
This damage is cumulative so pulling a sensor a couple of times a month
for inspection or cleaning, a five minute job, could cause the sensor
to fail in as little as six months
Q8.) What is the difference between fluorescence
and luminescence? BACK TO TOP
While the two methods are similar, there is one key difference. Fluorescence,
which is the method used by Insite, is the measurement of the immediate
reaction of a material in respone to an excitation energy source. Luminescence
is the the measurement of the time it takes the material to recover after
the excitation energy source is removed. This method is the one in use
by all other manufacturers currently marketing “optical” type
DO systems.
Q9.) What is the accuracy of the Insite fluorescence
DO system? BACK TO TOP
+/- 0.05 ppm
Q10.) What is the minimum flow required for
the sensor to properly measure DO? BACK
TO TOP
Unlike Clark type cells with a membrane, the Insite DO sensors do not
require any flow. Clark type DO sensors actually consume oxygen to make
the measurement so a new supply of oxygen must be continuously provided.
The Insite sensor does not consume anything so no flow is needed to obtain
a correct reading.
Q11.) Does the measuring area of the sensor
need to remain moist, even when not in actual use? BACK
TO TOP
No, the sensor can periodically become completely dry with no loss of
accuracy, response time, or calibration. No soaking caps are required
for storing the sensor.
Q12.) If the actual measuring area
of the sensor is physically damaged, can this be repaired? BACK
TO TOP
In the very unlikely event that the measuring surface is damaged it can
be repaired. The cost of this repair is around $200.00. We recently had
a customer put a system (analyzer and sensor) in the back of a golf cart
to take to the shop. The sensor did not quite make it into the cart and
was left hanging out the back on the ground. The sensor was then dragged
the entire trip back to the shop, a couple of thousand feet. The sensor
looked like it was put under a grinder but it still worked. This says
something about how hard it is to physically damage the sensor.
Q13.) Can the Insite DO systems be used in
very low oxygen environments, such as anoxic and anaerobic zones?
BACK TO TOP
Yes, the Insite method of measuring DO is very accurate at extremely low
ranges. There have been tests in which the unit performed very well in
the 0.03ppm to 0.08ppm range.
Q14.) What are the main reasons to
monitor and control DO in biological reactors?
BACK TO TOP
While there are almost as many specific reasons as there are treatment
facilities, they generally fall into one of three general areas.
1- Reduce the amount of power required to run the blowers.
This in turn will significantly reduce plant costs.
In general, up to 70% of a facility’s power consumption is for aeration.
An industry association estimates that the average treatment
facility could save up to 30% by automatically controlling
aeration.
2- Optimize the conditions in the basins to achieve
the correct microbiological mix. This is especially important
in BNR facilities.
3- Reduce the overall maintenance required to operate
the biological reactors. |